XCPMain Title Xceed POCAT menu XCPMain XCPMain XceedPro Main Menu XCPMain Button Xceed POCAT ToSpot SpecialNext Begin XceedPro Training XCPMain Button Xceed POCAT menu Review Review Specific Topics XCP Title Xceed POCAT menu POCCTPostTest XCPMain XceedPro Main Menu XCP Button Xceed POCAT ToSpot SpecialNext Begin XceedPro Training XCP Button Xceed POCAT menu Review Review Specific Topics XCP Button Xceed POCAT ToSpot SpecialTest Competency Test XCP Title Xceed POCAT menu Review XCPMain Review XceedPro Topics XCP Button Xceed POCAT ToSpot Parts Parts of XceedPro XCP Button Xceed POCAT ToSpot Operation Operation XCP Button Xceed POCAT ToSpot Trouble Troubleshooting XCP Button Xceed POCAT ToSpot Review-Compare Comparison of XceedPro and Precision PCx XCP Title Xceed ACT-1 topic Intro XceedPro Training XCP Body Xceed ACT-2 Welcome to the Precision XceedPro Blood Glucose Monitoring System Operator Training Video. XCP Body Xceed ACT-3 Abbott Diabetes Care created this guide to describe how to collect blood samples and perform patient tests and control tests using the Precision XceedPro System. XCP Body Xceed NoSound signin Signin POCCT Signin XCP Title Xceed POCAT menu Operation Review Operation XCP Button Xceed POCAT ToSpot Review-Sample Sample Collection XCP Button Xceed POCAT ToSpot Review-Capillary Capillary Blood Samples segment XCP Button Xceed POCAT ToSpot Review-Arterial Arterial Blood Samples XCP Button Xceed POCAT ToSpot Review-Venous Venous Blood Samples Section XCP Button Xceed POCAT ToSpot Review-PatientTest Patient Test XCP Button Xceed POCAT ToSpot Review-PatientID Patient ID Confirmation XCP Button Xceed POCAT ToSpot Review-Control Control Test XCP Button Xceed POCAT ToSpot Review-DataReview Data Review XCP Title Xceed POCAT menu Trouble Review Operation XCP Button Xceed POCAT ToSpot Review-TS-PatientTest Troubleshooting Patient Test XCP Button Xceed POCAT ToSpot Review-TS-Control Troubleshooting Control Test XCP Title Xceed POCAT menu Parts Review Parts of XceedPro XCP Button Xceed POCAT ToSpot Review-Keypad Keypad XCP Button Xceed POCAT ToSpot Review-Barcode Barcode Scanner XCP Button Xceed POCAT ToSpot Review-DataPort Data Port XCP Button Xceed POCAT ToSpot Review-Battery Battery Compartment XCP Button Xceed POCAT ToSpot Review-Port Port Protector XCP Button Xceed POCAT ToSpot Review-TestStrip Test Strip XCP Button Xceed POCAT ToSpot Review-Cleaning Cleaning XCP Button Xceed POCAT ToSpot Review-Isolation Isolation Bags XCP Admin Keypad admin Admin 933 POCCT Administration Menu XCP Special special SpecialNext SpecialNext XCP Special special SpecialTest SpecialTest XCP Title PatientTest video PatientTest 867 Patient Test XCP Body 3400 To Perform a Patient Test: XCP Body 9000 Press the On-Off button to turn on the monitor. XCP Body 14000 Select 1 - Patient Test. XCP Body 18800 Scan or enter the Operator ID. XCP Body 25867 "To scan an ID, hold the monitor 3 to 12 inches from the barcode." XCP Body 31533 Scan or enter the Patient ID. XCP Body 35400 "Now, Scan the test strip barcode. " XCP Body 48200 "If you are manually entering the test strip lot number, use the keypad to enter the number located on the barcode label, then press enter." XCP Body 52600 "If prompted, select the sample type." XCP Body 60867 Open the foil test strip packet at the notch and tear down to remove the test strip. XCP Body 69000 Insert the strip with the contact bars facing the monitor. XCP Body 74800 Apply the blood sample to the target area of the test strip. XCP Body 82267 "Remember, do not allow blood to run down the test strip into the port." XCP Body 92467 The monitor displays the test result after the countdown is completed. XCP Body 98733 "You may see additional post-testing prompts, such as Comment Code or Free Text entry. " XCP Body 100800 Scan or enter this information if prompted. XCP Body 110600 "Once the result is displayed, remove the test strip from the monitor and discard it as required by your facility's biohazard policy." XCP Body 125067 "If a test result is outside the defined Action Range, the result will appear with a triangle pointing up for ""above"" or pointing down for ""below"" the range. " XCP Body 136467 "If a test result is outside of the defined Critical Range, the result will appear with a greater than or less than symbol. " XCP Body 142533 Refer to your institution's policy for handling out-of-range test results. XCP QuesSetup Clipboard QuesSetup topic Now I would like to ask you some questions to make sure I have explained this clearly. XCP Question POCAT-Q1 question Is it OK to let blood run down the test strip into the port? XCP Touch Touch Please touch your answer. XCP Answer Yes PatientTest Yes XCP Answer-R No SpecialNext No XCP Answer NotSure PatientTest Not Sure XCP Title PatientID video PatientID 467 Patient ID Confirmation XCP Body 9933 When running a Patient Test you may receive an additional prompt to confirm the patient's ID number. XCP Body 15867 This prompt will appear after the initial Patient ID entry screen. XCP Body 21933 To Perform a Patient Test with a patient ID confirmation screen enabled:: XCP Body 27000 Press the On-Off button to turn on the monitor. XCP Body 31800 Select 1 - Patient Test. XCP Body 37133 Scan or enter the Operator ID. XCP Body 40867 Scan or enter the Patient ID. XCP Body 45333 "After the initial entry, you will be prompted to confirm the Patient ID. " XCP Body 52333 You may see one of the following three screens: XCP Body 56400 Repeat Manual ID Entry Screen XCP Body 62333 Re-enter the Patient ID using the Keypad to confirm the patient's ID. XCP Body 65000 Confirm Patient ID Entry Screen XCP Body 73400 Press 2 to confirm the Patient ID and continue testing or 1 to re-enter the ID. XCP Body 77400 Confirm Patient's Year of Birth Entry Screen XCP Body 84733 Enter the patient's year of birth to confirm the Patient ID or press clear to re-enter the ID. XCP Body 95333 "If you are prompted with a message Patient Data Not Found, the monitor either cannot confirm the Patient ID entered or the patient ID entered is not on the monitor's patient list. " XCP Body 99267 Press 1 to Re-Enter and confirm the patient ID. XCP Body 115533 Once you have successfully confirmed the patient's ID you will be prompted to Scan the test strip barcode. XCP Body 119933 "If prompted, select the sample type." XCP Body 130667 Open the foil test strip packet at the notch and tear down to remove the test strip. XCP Body 137200 Insert the strip with the contact bars facing the monitor. XCP Body 145000 Apply the blood sample to the target area of the test strip. XCP Body 152867 "Remember, do not allow blood to run down the test strip into the port." XCP Body 161733 The monitor displays the test result after the countdown is completed. XCP Body 170200 "You may see additional post-testing prompts, such as Comment Code or Free Text entry. " XCP Body 177000 Scan or enter this information if prompted. XCP Body 183000 "Once the result is displayed, remove the test strip from the monitor and discard it as required by your facility's biohazard policy." XCP QuesSetup Clipboard QuesSetup topic Now I would like to ask you some questions to make sure I have explained this clearly. XCP Question POCAT-Q2 question Do you have to enter both an Operator ID and a Patient ID to start a test? XCP Touch Touch Please touch your answer. XCP Answer-R Yes SpecialNext Yes XCP Answer No PatientID No XCP Answer NotSure PatientID Not Sure XCP Title Xceed NoSound topic Begin XceedPro Training XCP Body PCXCT-1 NoSound This training takes about 6 minutes and has questions to check your understanding. XCP Body PCXCT-1 NoSound After you complete the training you will be asked 5 randomly chosen questions to certify your competency to use the XceedPro. XCP Title Xceed Parts-1 topic Parts of the XceedPro XCP Title Keypad video Keypad 933 Keypad XCP Body 11933 Use the keypad to enter identification numbers and letters or to select an option that appears on the display. XCP Body 21933 The keypad contains a 10 digit telephone-style keypad with keys for the numbers zero through 9. XCP Body 30200 Buttons 2 through 9 also have letters below the number for entering alphabetical characters. XCP Body 36333 "To enter upper case characters, press the alphanumeric button. " XCP Body 42467 Press this button twice to enter lowercase characters. XCP Body 47667 And press again to return to entering numbers. XCP Body SpecialNext 62500 "The keypad also has six special keys - a scan, on/off, backlight/alphanumeric, clear, menu, and enter." XCP Title Barcode video Barcode 1600 Barcode Scanner XCP Body 12533 "To Scan, hold the monitor 3 to 12 inches from the barcode - at an angle from 30 to 135 degrees to the barcode." XCP Body 22000 Press and hold the scan button until a visible red beam is emitted by the monitor. XCP Body 30400 "Slowly move the monitor, if necessary, so the red beam is directly over the barcode. " XCP Body SpecialNext 33533 The monitor will beep when the scan is complete. XCP Title DataPort video DataPort 933 Data Port XCP Body 7533 The Data Port is located on the bottom of the monitor. XCP Body SpecialNext 18033 This port enables the monitor to automatically transfer the data through a cable or docking station to the data management system. XCP Title Battery video Battery 333 Battery Compartment XCP Body 9933 The monitor is powered by two AA batteries that go in the rear compartment. XCP Body 15933 Two nickel-metal hydride (NiMH) rechargeable batteries may also be used. XCP Body 22200 "To prolong the battery life, the monitor should be turned off when not in use. " XCP Body SpecialNext 29533 Note that stored data is not affected if the batteries are removed. XCP Title Port video Port 333 Port Protector XCP Body 11400 The port protector helps keep liquid from entering the test strip port and damaging the monitor. XCP Body 17200 Replace the protector if it comes in contact with blood or control solution. XCP Body 22933 "When replacement is necessary, lift the protector from the left or right edge and gently pull away. " XCP Body 31067 Then rest the flat bottom of the new protector on the test strip port and press gently into place on both sides. XCP Body SpecialNext 35733 "After cleaning, dry the protector thoroughly." XCP Title TestStrip video TestStrip 467 Test Strip XCP Body 9933 The Precision XceedPro monitor is compatible with Precision PCx Plus Blood Glucose Test Strips. XCP Body 17533 Precision PCx Plus Test Strips come in a gold foil packet with a barcode label. XCP Body 29067 "The individual foil packet ensures the integrity of each test strip by protecting it from exposure to air, moisture, and accidental contamination. " XCP Body 40533 "The bar code label holds information about the test strip, including the lot number, calibration, expiration date, and expected control solution ranges. " XCP Body 48467 One quick scan sends all this information to the monitor at the time of the test. XCP Body 56667 "To open the foil packet, grasp it at the notch, tear down, and remove the test strip." XCP Body 65067 The test strip is designed for either top-fill or end-fill sample application. XCP Body SpecialNext 76533 "Insert the contact bars into the monitor first, then apply blood or control solution to the target area at the other end of the test strip" XCP Title Cleaning video Cleaning 467 Cleaning XCP Body 11400 Clean the monitor by wiping it with a damp cloth or sponge moistened with a mild detergent. XCP Body 18267 Please refer to the Operator's Manual for a list of approved cleaning solutions. XCP Body SpecialNext 26533 Do not spray cleaner directly onto the monitor or allow liquid to drip into the strip port. XCP Title Isolation video Isolation 333 Isolation Bags XCP Body 5800 Disposable plastic isolation bags are available to provide a partial barrier between the monitor and patients in isolation. XCP Body SpecialNext 12533 "Insert the monitor into the bag so the strip port aligns with the hole, and seal the opposite end." XCP Title Sample video Sample 733 Sample Collection XCP Body 7867 The following are Operating Guidelines for tests using all sample types. XCP Body 15333 Use only Precision PCx Plus Test Strips with the Precision XceedPro Monitor. XCP Body 25867 "For specific directions on storage and use of the test strips, refer to the package insert in the test strip box." XCP Body 35867 Do not use the test strips beyond the expiration date printed on the barcode label and outer box. XCP Body 41867 "Do not use wet, bent, scratched or damaged test strips." XCP Body 47467 Always use test strips immediately after opening the foil packet. XCP Body 54000 Do not use test strips if the foil has a puncture or tear. XCP Body 59467 "Use the test strip only once, and then discard it." XCP Body 69867 Use only control solutions listed on the test strip package insert to verify the performance of the Precision XceedPro System. XCP Body 80333 Do not scan a packet's barcode and use a test strip from another packet. This could generate incorrect test results. XCP Body 86200 Make sure the monitor and test strips are at room temperature. XCP Body 96467 "If the monitor is moved to a warmer or cooler area, allow the monitor to reach the new room temperature." XCP Body 106867 Always wear gloves. And follow your facility's biohazard safety policies and procedures when performing tests involving patient blood samples. XCP Body 114867 Fill the entire target area of the test strip with the blood sample. XCP Body 126467 "The test results will not be affected if the target area has been briefly touched with the patient's finger, a capillary tube or pipette." XCP Body 137333 "After the blood has been applied to the test strip and the test countdown begins, do not remove or disturb the test strip." XCP Body 145467 "If the test fails to start, quickly apply a second drop of blood to the target area. " XCP Body 155933 "If the test fails to start after the second drop is applied, discard the used test strip and repeat the test. " XCP Body 165000 The allowable time limit to add a second drop of blood can be found in the test strip package insert. XCP Title Capillary video Capillary SpecialNext 400 Capillary Blood Samples segment XCP Body 5867 The following guidelines apply to tests using Capillary Blood Samples XCP Body 9733 Use only fresh whole blood samples. XCP Body 20000 Clean and completely dry the puncture site. Make sure the sampling site is clean and dry before lancing. XCP Body 30133 Collect the capillary blood using a lancing device and an appropriate technique. XCP Body 42800 Apply a drop of blood to the target area at the end of the test strip. Allow the entire target area to fill with blood. XCP Body SpecialNext 55033 "If necessary, the blood can be collected in a capillary tube coated with heparin or EDTA then applied to the test strip within 30 minutes of collection." XCP Title Arterial video Arterial 733 Arterial Blood Samples XCP Body 6333 The following guidelines apply to tests using Arterial Blood Samples: XCP Body 10600 Use only fresh whole blood samples. XCP Body 18333 Clear the arterial line before drawing a blood sample into the heparinized syringe. XCP Body 22933 Use the sample within 30 minutes of collection. XCP Body 40733 "Immediately before applying the sample to the target area, mix the sample well by rolling the syringe between outstretched palms of both hands for at least 5 seconds; invert the syringe and repeat mixing." XCP Body 46000 Allow a drop of blood to form at the tip of the syringe. XCP Body 59533 Apply a drop of blood to the target area at the end of the test strip. Allow the entire target area to fill with blood. XCP Body SpecialNext 67033 The syringe can briefly touch the test strip without affecting the test result. XCP Title Venous video Venous 867 Venous Blood Samples Section XCP Body 6467 The following guidelines apply to tests using Venous Blood Samples. XCP Body 10333 Use only fresh whole blood samples. XCP Body 18067 Collect the venous blood sample in a collection tube containing heparin or EDTA. XCP Body 26067 Make sure the test tube is filled to the volume specified. Do not under-fill. XCP Body 32133 Do not use collection tubes that contain fluoride or oxalate. XCP Body 44400 "If the blood is collected from an intravenous line, clear the line before drawing the sample into a syringe that contains heparin or EDTA." XCP Body 48933 Use the sample within 30 minutes of collection. XCP Body 57667 Immediately invert the tube with the sample several times before removing the sample. XCP Body 66467 Use a disposable transfer pipette to obtain a sample from the center of the collection tube. XCP Body 72400 Apply the drop of blood to the end of the test strip. XCP Body 90067 "If you are prompted to 'Enter Sample Type' when using the Precision PCx Plus Test Strip, use venous mode only for venous samples that have been in capped tubes and have not been mixed with air. " XCP Body SpecialNext 96033 Use capillary mode for all other samples. XCP Title Control video Control 1267 Control Test XCP Body 3867 To Perform a Control Test XCP Body 9467 Press the on-Off button to turn on the monitor. XCP Body 13667 Select Two - Control Test. XCP Body 18067 Scan or enter the Operator ID XCP Body 23400 Scan or enter the control Solution Lot number. XCP Body 29133 Scan or enter the test strip lot number. XCP Body 37733 "Open the foil test strip packet at the notch, tear down, and remove the test strip." XCP Body 44533 Insert the Strip with the contact bars facing the monitor. XCP Body 49733 Wipe the control solution nozzle with a clean gauze or tissue. XCP Body 57867 Then apply a small drop of solution to the test strip target area. XCP Body 66333 Remember: Do not allow control solution to run down the test strip into the port XCP Body 74067 Do not lift the monitor to view the display until after the test strip has been removed. XCP Body 83467 The monitor displays the test result after the countdown is completed. XCP Body 89800 Results may be displayed as a number or as Pass/Fail. XCP Body 101333 "If the control result is a number, an 'X' appears in front of a result that is out of range." XCP Body SpecialNext 106833 "If prompted, enter a comment code." XCP Title DataReview video DataReview 667 Data Review XCP Body 5733 There are two ways to review test data: XCP Body 16067 Review a patient's result history after completing test or view test results through the Data Review menu option. XCP Body 24667 Upon completion of a patient test you will be prompted with an option to view Patient History. XCP Body 32200 Select 2 - Patient History to view the test result history of the patient just tested. XCP Body 42533 "To view test result through the Data Review menu option, select the Menu button on the keypad. " XCP Body 47733 Select 1 - Data Review. Scan or enter your Operator ID if prompted. XCP Body SpecialNext 61500 Then select the desired data review option from the sub-menu. XCP Title Compare video Compare 600 Comparison Chapter XCP Body 11933 Users of the Abbott Diabetes Care Precision XceedPro may also be familiar with the Abbott Diabetes Care Precision PCx Blood Glucose Monitoring System. XCP Body 23800 The following section provides a side by side comparison of the Precision XceedPro and the Precision PCx Blood Glucose Monitors. XCP Body 29933 The Precision XceedPro monitor is similar to Abbott's Precision PCx Blood Glucose monitor. XCP Body 35400 It builds on that proven success - providing all the same features. XCP Body 40867 "You see the same ergonomic design, for hand-held comfort." XCP Body 48333 "In both, the Test Strip Port is located at the top of the monitor." XCP Body 54267 "Both have a large, clear LCD screen - for accurate viewing." XCP Body 59400 The scan button is at the top of the keypad. XCP Body 64867 The Data Port is on the bottom of the monitor. XCP Body 70467 The battery compartment is on back of the monitor. XCP Body 75800 There's an integrated barcode scanner. XCP Body 86000 You can even use the familiar gold foiled Precision PCx Plus Blood Glucose Test Strips with the Precision XceedPro monitor. XCP Body 93533 And the method of applying blood samples and test solutions is also the same. XCP Body 99267 All the same great features - with a few new enhancements. XCP Body 112333 A Port Protector has been added to the test strip port to help keep liquid from entering the test strip port and damaging the monitor. XCP Body 119400 An alphanumeric button has been added to the keypad to support additional data entry options. XCP Body SpecialNext 130033 "The Clear, Menu and Enter keys were relocated to the top of the keypad so it's easier to use." XCP Title Keypad video Review-Keypad 933 Keypad XCP Body 11933 Use the keypad to enter identification numbers and letters or to select an option that appears on the display. XCP Body 21933 The keypad contains a 10 digit telephone-style keypad with keys for the numbers zero through 9. XCP Body 30200 Buttons 2 through 9 also have letters below the number for entering alphabetical characters. XCP Body 36333 "To enter upper case characters, press the alphanumeric button. " XCP Body 42467 Press this button twice to enter lowercase characters. XCP Body 47667 And press again to return to entering numbers. XCP Body Parts 62500 "The keypad also has six special keys - a scan, on/off, backlight/alphanumeric, clear, menu, and enter." XCP Title Barcode video Review-Barcode 1600 Barcode Scanner XCP Body 12533 "To Scan, hold the monitor 3 to 12 inches from the barcode - at an angle from 30 to 135 degrees to the barcode." XCP Body 22000 Press and hold the scan button until a visible red beam is emitted by the monitor. XCP Body 30400 "Slowly move the monitor, if necessary, so the red beam is directly over the barcode. " XCP Body Parts 33533 The monitor will beep when the scan is complete. XCP Title DataPort video Review-DataPort 933 Data Port XCP Body 7533 The Data Port is located on the bottom of the monitor. XCP Body Parts 18033 This port enables the monitor to automatically transfer the data through a cable or docking station to the data management system. XCP Title Battery video Review-Battery 333 Battery Compartment XCP Body 9933 The monitor is powered by two AA batteries that go in the rear compartment. XCP Body 15933 Two nickel-metal hydride (NiMH) rechargeable batteries may also be used. XCP Body 22200 "To prolong the battery life, the monitor should be turned off when not in use. " XCP Body Parts 29533 Note that stored data is not affected if the batteries are removed. XCP Title Port video Review-Port 333 Port Protector XCP Body 11400 The port protector helps keep liquid from entering the test strip port and damaging the monitor. XCP Body 17200 Replace the protector if it comes in contact with blood or control solution. XCP Body 22933 "When replacement is necessary, lift the protector from the left or right edge and gently pull away. " XCP Body 31067 Then rest the flat bottom of the new protector on the test strip port and press gently into place on both sides. XCP Body Parts 35733 "After cleaning, dry the protector thoroughly." XCP Title TestStrip video Review-TestStrip 467 Test Strip XCP Body 9933 The Precision XceedPro monitor is compatible with Precision PCx Plus Blood Glucose Test Strips. XCP Body 17533 Precision PCx Plus Test Strips come in a gold foil packet with a barcode label. XCP Body 29067 "The individual foil packet ensures the integrity of each test strip by protecting it from exposure to air, moisture, and accidental contamination. " XCP Body 40533 "The bar code label holds information about the test strip, including the lot number, calibration, expiration date, and expected control solution ranges. " XCP Body 48467 One quick scan sends all this information to the monitor at the time of the test. XCP Body 56667 "To open the foil packet, grasp it at the notch, tear down, and remove the test strip." XCP Body 65067 The test strip is designed for either top-fill or end-fill sample application. XCP Body Parts 76533 "Insert the contact bars into the monitor first, then apply blood or control solution to the target area at the other end of the test strip" XCP Title Cleaning video Review-Cleaning 467 Cleaning XCP Body 11400 Clean the monitor by wiping it with a damp cloth or sponge moistened with a mild detergent. XCP Body 18267 Please refer to the Operator's Manual for a list of approved cleaning solutions. XCP Body Parts 26533 Do not spray cleaner directly onto the monitor or allow liquid to drip into the strip port. XCP Title Isolation video Review-Isolation 333 Isolation Bags XCP Body 5800 Disposable plastic isolation bags are available to provide a partial barrier between the monitor and patients in isolation. XCP Body Parts 12533 "Insert the monitor into the bag so the strip port aligns with the hole, and seal the opposite end." XCP Title Sample video Review-Sample 733 Sample Collection XCP Body 7867 The following are Operating Guidelines for tests using all sample types. XCP Body 15333 Use only Precision PCx Plus Test Strips with the Precision XceedPro Monitor. XCP Body 25867 "For specific directions on storage and use of the test strips, refer to the package insert in the test strip box." XCP Body 35867 Do not use the test strips beyond the expiration date printed on the barcode label and outer box. XCP Body 41867 "Do not use wet, bent, scratched or damaged test strips." XCP Body 47467 Always use test strips immediately after opening the foil packet. XCP Body 54000 Do not use test strips if the foil has a puncture or tear. XCP Body 59467 "Use the test strip only once, and then discard it." XCP Body 69867 Use only control solutions listed on the test strip package insert to verify the performance of the Precision XceedPro System. XCP Body 80333 Do not scan a packet's barcode and use a test strip from another packet. This could generate incorrect test results. XCP Body 86200 Make sure the monitor and test strips are at room temperature. XCP Body 96467 "If the monitor is moved to a warmer or cooler area, allow the monitor to reach the new room temperature." XCP Body 106867 Always wear gloves. And follow your facility's biohazard safety policies and procedures when performing tests involving patient blood samples. XCP Body 114867 Fill the entire target area of the test strip with the blood sample. XCP Body 126467 "The test results will not be affected if the target area has been briefly touched with the patient's finger, a capillary tube or pipette." XCP Body 137333 "After the blood has been applied to the test strip and the test countdown begins, do not remove or disturb the test strip." XCP Body 145467 "If the test fails to start, quickly apply a second drop of blood to the target area. " XCP Body 155933 "If the test fails to start after the second drop is applied, discard the used test strip and repeat the test. " XCP Body Operation 165000 The allowable time limit to add a second drop of blood can be found in the test strip package insert. XCP Title Capillary video Review-Capillary 400 Capillary Blood Samples segment XCP Body 5867 The following guidelines apply to tests using Capillary Blood Samples XCP Body 9733 Use only fresh whole blood samples. XCP Body 20000 Clean and completely dry the puncture site. Make sure the sampling site is clean and dry before lancing. XCP Body 30133 Collect the capillary blood using a lancing device and an appropriate technique. XCP Body 42800 Apply a drop of blood to the target area at the end of the test strip. Allow the entire target area to fill with blood. XCP Body Operation 55033 "If necessary, the blood can be collected in a capillary tube coated with heparin or EDTA then applied to the test strip within 30 minutes of collection." XCP Title Arterial video Review-Arterial 733 Arterial Blood Samples XCP Body 6333 The following guidelines apply to tests using Arterial Blood Samples: XCP Body 10600 Use only fresh whole blood samples. XCP Body 18333 Clear the arterial line before drawing a blood sample into the heparinized syringe. XCP Body 22933 Use the sample within 30 minutes of collection. XCP Body 40733 "Immediately before applying the sample to the target area, mix the sample well by rolling the syringe between outstretched palms of both hands for at least 5 seconds; invert the syringe and repeat mixing." XCP Body 46000 Allow a drop of blood to form at the tip of the syringe. XCP Body 59533 Apply a drop of blood to the target area at the end of the test strip. Allow the entire target area to fill with blood. XCP Body Operation 67033 The syringe can briefly touch the test strip without affecting the test result. XCP Title Venous video Review-Venous 867 Venous Blood Samples Section XCP Body 6467 The following guidelines apply to tests using Venous Blood Samples. XCP Body 10333 Use only fresh whole blood samples. XCP Body 18067 Collect the venous blood sample in a collection tube containing heparin or EDTA. XCP Body 26067 Make sure the test tube is filled to the volume specified. Do not under-fill. XCP Body 32133 Do not use collection tubes that contain fluoride or oxalate. XCP Body 44400 "If the blood is collected from an intravenous line, clear the line before drawing the sample into a syringe that contains heparin or EDTA." XCP Body 48933 Use the sample within 30 minutes of collection. XCP Body 57667 Immediately invert the tube with the sample several times before removing the sample. XCP Body 66467 Use a disposable transfer pipette to obtain a sample from the center of the collection tube. XCP Body 72400 Apply the drop of blood to the end of the test strip. XCP Body 90067 "If you are prompted to 'Enter Sample Type' when using the Precision PCx Plus Test Strip, use venous mode only for venous samples that have been in capped tubes and have not been mixed with air. " XCP Body Operation 96033 Use capillary mode for all other samples. XCP Title PatientTest video Review-PatientTest 867 Patient Test XCP Body 3400 To Perform a Patient Test: XCP Body 9000 Press the On-Off button to turn on the monitor. XCP Body 14000 Select 1 - Patient Test. XCP Body 18800 Scan or enter the Operator ID. XCP Body 25867 "To scan an ID, hold the monitor 3 to 12 inches from the barcode." XCP Body 31533 Scan or enter the Patient ID. XCP Body 35400 "Now, Scan the test strip barcode. " XCP Body 48200 "If you are manually entering the test strip lot number, use the keypad to enter the number located on the barcode label, then press enter." XCP Body 52600 "If prompted, select the sample type." XCP Body 60867 Open the foil test strip packet at the notch and tear down to remove the test strip. XCP Body 69000 Insert the strip with the contact bars facing the monitor. XCP Body 74800 Apply the blood sample to the target area of the test strip. XCP Body 82267 "Remember, do not allow blood to run down the test strip into the port." XCP Body 92467 The monitor displays the test result after the countdown is completed. XCP Body 98733 "You may see additional post-testing prompts, such as Comment Code or Free Text entry. " XCP Body 100800 Scan or enter this information if prompted. XCP Body 110600 "Once the result is displayed, remove the test strip from the monitor and discard it as required by your facility's biohazard policy." XCP Body 125067 "If a test result is outside the defined Action Range, the result will appear with a triangle pointing up for ""above"" or pointing down for ""below"" the range. " XCP Body 136467 "If a test result is outside of the defined Critical Range, the result will appear with a greater than or less than symbol. " XCP Body Operation 142533 Refer to your institution's policy for handling out-of-range test results. XCP Title PatientID video Review-PatientID 467 Patient ID Confirmation XCP Body 9933 When running a Patient Test you may receive an additional prompt to confirm the patient's ID number. XCP Body 15867 This prompt will appear after the initial Patient ID entry screen. XCP Body 21933 To Perform a Patient Test with a patient ID confirmation screen enabled:: XCP Body 27000 Press the On-Off button to turn on the monitor. XCP Body 31800 Select 1 - Patient Test. XCP Body 37133 Scan or enter the Operator ID. XCP Body 40867 Scan or enter the Patient ID. XCP Body 45333 "After the initial entry, you will be prompted to confirm the Patient ID. " XCP Body 52333 You may see one of the following three screens: XCP Body 56400 Repeat Manual ID Entry Screen XCP Body 62333 Re-enter the Patient ID using the Keypad to confirm the patient's ID. XCP Body 65000 Confirm Patient ID Entry Screen XCP Body 73400 Press 2 to confirm the Patient ID and continue testing or 1 to re-enter the ID. XCP Body 77400 Confirm Patient's Year of Birth Entry Screen XCP Body 84733 Enter the patient's year of birth to confirm the Patient ID or press clear to re-enter the ID. XCP Body 95333 "If you are prompted with a message Patient Data Not Found, the monitor either cannot confirm the Patient ID entered or the patient ID entered is not on the monitor's patient list. " XCP Body 99267 Press 1 to Re-Enter and confirm the patient ID. XCP Body 115533 Once you have successfully confirmed the patient's ID you will be prompted to Scan the test strip barcode. XCP Body 119933 "If prompted, select the sample type." XCP Body 130667 Open the foil test strip packet at the notch and tear down to remove the test strip. XCP Body 137200 Insert the strip with the contact bars facing the monitor. XCP Body 145000 Apply the blood sample to the target area of the test strip. XCP Body 152867 "Remember, do not allow blood to run down the test strip into the port." XCP Body 161733 The monitor displays the test result after the countdown is completed. XCP Body 170200 "You may see additional post-testing prompts, such as Comment Code or Free Text entry. " XCP Body 177000 Scan or enter this information if prompted. XCP Body Operation 183000 "Once the result is displayed, remove the test strip from the monitor and discard it as required by your facility's biohazard policy." XCP Title Control video Review-Control 1267 Control Test XCP Body 3867 To Perform a Control Test XCP Body 9467 Press the on-Off button to turn on the monitor. XCP Body 13667 Select Two - Control Test. XCP Body 18067 Scan or enter the Operator ID XCP Body 23400 Scan or enter the control Solution Lot number. XCP Body 29133 Scan or enter the test strip lot number. XCP Body 37733 "Open the foil test strip packet at the notch, tear down, and remove the test strip." XCP Body 44533 Insert the Strip with the contact bars facing the monitor. XCP Body 49733 Wipe the control solution nozzle with a clean gauze or tissue. XCP Body 57867 Then apply a small drop of solution to the test strip target area. XCP Body 66333 Remember: Do not allow control solution to run down the test strip into the port XCP Body 74067 Do not lift the monitor to view the display until after the test strip has been removed. XCP Body 83467 The monitor displays the test result after the countdown is completed. XCP Body 89800 Results may be displayed as a number or as Pass/Fail. XCP Body 101333 "If the control result is a number, an 'X' appears in front of a result that is out of range." XCP Body Operation 106833 "If prompted, enter a comment code." XCP Title DataReview video Review-DataReview 667 Data Review XCP Body 5733 There are two ways to review test data: XCP Body 16067 Review a patient's result history after completing test or view test results through the Data Review menu option. XCP Body 24667 Upon completion of a patient test you will be prompted with an option to view Patient History. XCP Body 32200 Select 2 - Patient History to view the test result history of the patient just tested. XCP Body 42533 "To view test result through the Data Review menu option, select the Menu button on the keypad. " XCP Body 47733 Select 1 - Data Review. Scan or enter your Operator ID if prompted. XCP Body Operation 61500 Then select the desired data review option from the sub-menu. XCP Title TS-PatientTest video Review-TS-PatientTest 1686 Troubleshooting Patient Test XCP Body 3148 Troubleshooting Patient Test Results XCP Body 8164 Important: If the blood glucose result appears to be inconsistent - either lower or higher than expected - there may be a problem with the test strip. XCP Body 17255 Repeat the test using a new test strip. XCP Body 20650 Results that are incorrect may have serious medical consequences. XCP Body 30995 Consult the prescribing physician before making any changes to diabetes medication plans if either of the following occurs: XCP Body 36115 The blood glucose results are not consistent with the physical symptoms AND you have ruled out common errors in technique. XCP Body 45780 or The blood glucose result is less than 50 milligrams per deciliter (2.8 millimoles per liter) or greater than 300 milligrams per deciliter (16.7 millimoles per liter). XCP Body 50744 Possible reasons why glucose results may be higher than expected: XCP Body 55890 "The patient's hematocrit is lower than the acceptable limit for the test strips, as indicated on the test strip package insert." XCP Body 59756 Serum or plasma samples were used instead of whole blood. XCP Body 65215 Possible reasons why glucose results may be lower than expected: XCP Body 70623 "The patient's hematocrit is higher than the acceptable limit for the test strips, as indicated on the test strip package insert." XCP Body 75064 The patient is in a Hyperglycemic-hypersomolar state - with or without ketosis. XCP Body 77963 "The patient is experiencing severe dehydration, hypotension or shock." XCP Body 86636 There is water or alcohol remaining on the puncture site. XCP Body 91338 The venous or arterial whole blood sample was not tested within 30 minutes of collection. XCP Body Trouble 99132 "For more detailed information, please refer to the appropriate test strip package insert." XCP Title TS-Control video Review-TS-Control 901 Troubleshooting Control Test XCP Body 8509 "To Troubleshoot Out-of-Range Control test results, repeat the control test to make sure you meet the following conditions." XCP Body 16350 Eliminate any air bubbles in the control bottle's nozzle. XCP Body 25492 Wipe away excess solution from the bottle tip prior to testing. XCP Body 34101 Scan or enter the barcode from the packet of the test strip used. XCP Body 43810 Scan or enter the correct lot number for the control solution used. XCP Body 47080 Check the storage temperature of materials. XCP Body 52753 Control solutions and test strips must be stored at temperatures between 39 degrees and 86 degrees Fahrenheit or 4 degrees and 30 degrees Celsius. XCP Body 61361 Check that the temperature conditions in the room where the tests are being performed are within the acceptable operating range as specified in the test strip package insert. XCP Body 69636 Check that the bottles of control solution have not been open for more than 90 days. XCP Body 78579 Use a new test strip for each test. XCP Body 87888 Use only PCx Plus Test Strips. XCP Body 97231 "Use only MediSenseŽ, PrecisionŽ or Optium Control Solutions." XCP Body Trouble 110043 "If the above criteria are met and the test results are still out of range, please repeat the test using a new box of control solution and/or test strips." XCP Title Compare video Review-Compare 600 Comparison Chapter XCP Body 11933 Users of the Abbott Diabetes Care Precision XceedPro may also be familiar with the Abbott Diabetes Care Precision PCx Blood Glucose Monitoring System. XCP Body 23800 The following section provides a side by side comparison of the Precision XceedPro and the Precision PCx Blood Glucose Monitors. XCP Body 29933 The Precision XceedPro monitor is similar to Abbott's Precision PCx Blood Glucose monitor. XCP Body 35400 It builds on that proven success - providing all the same features. XCP Body 40867 "You see the same ergonomic design, for hand-held comfort." XCP Body 48333 "In both, the Test Strip Port is located at the top of the monitor." XCP Body 54267 "Both have a large, clear LCD screen - for accurate viewing." XCP Body 59400 The scan button is at the top of the keypad. XCP Body 64867 The Data Port is on the bottom of the monitor. XCP Body 70467 The battery compartment is on back of the monitor. XCP Body 75800 There's an integrated barcode scanner. XCP Body 86000 You can even use the familiar gold foiled Precision PCx Plus Blood Glucose Test Strips with the Precision XceedPro monitor. XCP Body 93533 And the method of applying blood samples and test solutions is also the same. XCP Body 99267 All the same great features - with a few new enhancements. XCP Body 112333 A Port Protector has been added to the test strip port to help keep liquid from entering the test strip port and damaging the monitor. XCP Body 119400 An alphanumeric button has been added to the keypad to support additional data entry options. XCP Body POCCTMain 130033 "The Clear, Menu and Enter keys were relocated to the top of the keypad so it's easier to use." XCP Title Clipboard ACT-4 topic Test SpecialTest XceedPro Competency Test ACT TestQuestion ACT-Q1 question CT-1 Control "A bottle of control solution is opened today, when will this bottle expire?" ACT Touch Touch Control Please touch your answer. ACT Answer ACT-Q1A1 SpecialTest Control The expiration date on the bottle ACT Answer ACT-Q1A2 SpecialTest Control 90 days after the bottle is opened ACT Answer ACT-Q1A3 SpecialTest Control The expiration date established by the laboratory ACT Answer-R ACT-Q1A4 SpecialTest Control " 90 days after the bottle is opened or the expiration date, whichever comes first" ACT TestQuestion ACT-Q2 question CT-2 PatientTest "By scanning the Test Strip, what information is captured?" ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q2A1 SpecialTest PatientTest Test Strip Lot Number ACT Answer ACT-Q2A2 SpecialTest PatientTest Test Strip Lot Expiration Date ACT Answer ACT-Q2A3 SpecialTest PatientTest Calibration Code ACT Answer ACT-Q2A4 SpecialTest PatientTest Quality Control Ranges ACT Answer-R ACT-Q2A5 SpecialTest PatientTest All of the Above ACT TestQuestion ACT-Q3 question CT-3 Compare Which feature is available on the the Precision XceedPro that was not on the Precision PCx? ACT Touch Touch Compare Please touch your answer. ACT Answer ACT-Q3A1 SpecialTest Compare Laser Barcode Scanner ACT Answer ACT-Q3A2 SpecialTest Compare Low Battery Warning ACT Answer-R ACT-Q3A5 SpecialTest Compare Port Protector ACT Answer ACT-Q3A3 SpecialTest Compare Clear and Menu Keys ACT Answer ACT-Q3A4 SpecialTest Compare Patient and Control Testing Prompts ACT TestQuestion ACT-Q4 question CT-4 PatientTest How many Patient Test Records can be stored in the memory of the Precision Xceed Pro meter? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q4A1 SpecialTest PatientTest 100 ACT Answer ACT-Q4A2 SpecialTest PatientTest 500 ACT Answer ACT-Q4A3 SpecialTest PatientTest 1000 ACT Answer ACT-Q4A4 SpecialTest PatientTest 2000 ACT Answer-R ACT-Q4A5 SpecialTest PatientTest 2500 ACT TestQuestion ACT-Q5 question CT-5 Sample How do you know if you have applied enough blood to the test strip of the Precision Xceed Pro? ACT Touch Touch Sample Please touch your answer. ACT Answer ACT-Q5A1 SpecialTest Sample You must check the strip visually. ACT Answer ACT-Q5A2 SpecialTest Sample You will not know if you have enough sample for the test. ACT Answer ACT-Q5A3 SpecialTest Sample You see the screen change to apply blood. ACT Answer ACT-Q5A4 SpecialTest Sample You will see the blood turn purple on the test strip. ACT Answer-R ACT-Q5A5 SpecialTest Sample You will hear an audible beep and the testing will automatically start. ACT TestQuestion ACT-Q6 question CT-6 Sample What are some reasons a blood glucose test may be higher than expected? ACT Touch Touch Sample Please touch your answer. ACT Answer ACT-Q6A1 SpecialTest Sample ACT Answer ACT-Q6A2 SpecialTest Sample ACT Answer ACT-Q6A3 SpecialTest Sample ACT Answer ACT-Q6A4 SpecialTest Sample ACT Answer-R ACT-Q6A5 SpecialTest Sample ACT TestQuestion ACT-Q7 question CT-7 PatientTest What are you expected to do when the Precision XceedPro prompts you to Scan or Enter Operator ID? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q7A1 SpecialTest PatientTest Scan the test strip barcode. ACT Answer ACT-Q7A2 SpecialTest PatientTest Type in the Patients Name. ACT Answer ACT-Q7A3 SpecialTest PatientTest Scan the Patients wrist band. ACT Answer-R ACT-Q7A5 SpecialTest PatientTest Scan or type in my ID number. ACT Answer ACT-Q7A4 SpecialTest PatientTest Enter in the calibration code. ACT TestQuestion ACT-Q8 question CT-8 Battery The Precision XceedPro meter will display a Battery Symbol indicating the batteries are running low. ACT Touch Touch Battery Please touch your answer. ACT Answer-R ACT-Q8A1 SpecialTest Battery TRUE ACT Answer ACT-Q8A2 SpecialTest Battery FALSE ACT TestQuestion ACT-Q9 question CT-9 Sample "Blood should never be applied to the Precision Xceed Test Strip until the strip is inserted into the meter and the display reads, Apply Sample." ACT Touch Touch Sample Please touch your answer. ACT Answer-R ACT-Q9A1 SpecialTest Sample TRUE ACT Answer ACT-Q9A2 SpecialTest Sample FALSE ACT TestQuestion ACT-Q10 question CT-10 Control Quality Control Testing should be performed before every patient test. ACT Touch Touch Control Please touch your answer. ACT Answer ACT-Q10A1 SpecialTest Control TRUE ACT Answer-R ACT-Q10A2 SpecialTest Control FALSE ACT TestQuestion ACT-Q11 question CT-11 Barcode What is the best way to scan a barcode using the Precision Xceed Pro? ACT Touch Touch Barcode Please touch your answer. ACT Answer ACT-Q11A1 SpecialTest Barcode Hold the Scan button and aim Precision scanner at a 90 degree angle to the barcode? ACT Answer ACT-Q11A2 SpecialTest Barcode "Press the scan button, release it and scan the barcode from paper." ACT Answer ACT-Q11A3 SpecialTest Barcode "Press and hold the enter button, while scanning the barcode." ACT Answer-R ACT-Q11A5 SpecialTest Barcode "3-6 inches from the barcode, slowly scan the barcode with the red beam until you hear a beep." ACT TestQuestion ACT-Q12 question CT-12 Compare The Precision Xceed ProŽ Plus blood glucose test strip is designed to allow which of the following: ACT Touch Touch Compare Please touch your answer. ACT Answer ACT-Q12A1 SpecialTest Compare Top Fill Sample application ACT Answer ACT-Q12A2 SpecialTest Compare End Fill sample application ACT Answer ACT-Q12A3 SpecialTest Compare Neither Top fill or End fill ACT Answer-R ACT-Q12A5 SpecialTest Compare Both Top Fill and End Fill sample application ACT TestQuestion ACT-Q13 question CT-13 Port "If blood or control solution comes in contact with the port protector, you should you replace port protector prior to testing." ACT Touch Touch Port Please touch your answer. ACT Answer-R ACT-Q13A1 SpecialTest Port TRUE ACT Answer ACT-Q13A2 SpecialTest Port FALSE ACT TestQuestion ACT-Q14 question CT-14 Keypad How do you backup one space while entering numbers on the keypad? ACT Touch Touch Keypad Please touch your answer. ACT Answer-R ACT-Q14A5 SpecialTest Keypad Press the CLEAR key ACT Answer ACT-Q14A1 SpecialTest Keypad Press the DELETE key ACT Answer ACT-Q14A2 SpecialTest Keypad Press the ENTER key ACT Answer ACT-Q14A3 SpecialTest Keypad Press the SCAN key ACT Answer ACT-Q14A4 SpecialTest Keypad Shut the meter OFF ACT TestQuestion ACT-Q15 question CT-15 PatientTest When can Patient test result history be recalled on the Precision XceedPro meter? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q15A1 SpecialTest PatientTest Before the test strip is scanned. ACT Answer ACT-Q15A2 SpecialTest PatientTest After the test strip is scanned ACT Answer ACT-Q15A3 SpecialTest PatientTest Patient test history cannot be recalled on the Precision XceedPro meter. ACT Answer-R ACT-Q15A5 SpecialTest PatientTest Either before the a test is run using Data Review or after a test is run using Patient History selections. ACT TestQuestion ACT-Q16 question CT-16 PatientTest Where can you find information to troubleshoot a patient or control test result that is out of range? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q16A1 SpecialTest PatientTest Opertor's Manual ACT Answer ACT-Q16A2 SpecialTest PatientTest Abbott Technical Support ACT Answer ACT-Q16A3 SpecialTest PatientTest Precision XceedPro Quick Reference Guide ACT Answer ACT-Q16A4 SpecialTest PatientTest Precision XceedPro Test Strip package insert ACT Answer-R ACT-Q16A5 SpecialTest PatientTest All of the above ACT TestQuestion ACT-Q17 question CT-17 Cleaning Is it OK to use alcohol and ammonia based cleaners for cleaning the Precision Xceed Pro monitor? ACT Touch Touch Cleaning Please touch your answer. ACT Answer-R ACT-Q17A1 SpecialTest Cleaning TRUE ACT Answer ACT-Q17A2 SpecialTest Cleaning FALSE ACT TestQuestion ACT-Q18 question CT-18 Isolation How can you use the Precision XceedPro meter with a patient on isolation? ACT Touch Touch Isolation Please touch your answer. ACT Answer ACT-Q18A1 SpecialTest Isolation Dedicate one Precision XceedPro meter to the isolation patient. ACT Answer ACT-Q18A2 SpecialTest Isolation Use a Protex isolation bag or cover the PXP meter using a bag prior to inserting strip ACT Answer ACT-Q18A3 SpecialTest Isolation Clean the surfaces of the Precision XceedPro Meter in between each patient ACT Answer-R ACT-Q18A5 SpecialTest Isolation All of the above. ACT TestQuestion ACT-Q19 question CT-19 Control "What is expected if you receive the message, "" Unexpected Level"" when running a Control test on the Precision XceedPro meter?" ACT Touch Touch Control Please touch your answer. ACT Answer ACT-Q19A1 SpecialTest Control Repeat the patient test ACT Answer ACT-Q19A2 SpecialTest Control Open a new control solution bottle. ACT Answer-R ACT-Q19A5 SpecialTest Control Select ReEnter Level and scan the level it is expecting. ACT Answer ACT-Q19A3 SpecialTest Control Enter in the lot number of the test strip. ACT Answer ACT-Q19A4 SpecialTest Control Scan the Patient's ID. ACT TestQuestion ACT-Q20 question CT-20 Patient ID Can you enter in letters instead of numbers in the Precision XceedPro? ACT Touch Touch Patient ID Please touch your answer. ACT Answer ACT-Q20A1 SpecialTest Patient ID "Yes, but you have to scan from a barcode." ACT Answer ACT-Q20A2 SpecialTest Patient ID No ACT Answer ACT-Q20A3 SpecialTest Patient ID Yes by using the Aa key only. ACT Answer-R ACT-Q20A5 SpecialTest Patient ID "Yes, by using the Aa key or scanning a barcode." ACT TestQuestion ACT-Q21 question CT-21 PatientTest What happens if you mistakenly hit the On/Off button on the Precision XceedPro during testing? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q21A1 SpecialTest PatientTest It stops the test and shuts off. ACT Answer ACT-Q21A2 SpecialTest PatientTest It displays a test error code ACT Answer ACT-Q21A3 SpecialTest PatientTest The test will continue and then it shuts off automatically when the test is complete ACT Answer-R ACT-Q21A5 SpecialTest PatientTest "Nothing, you must press and hold the On/Off key for the meter to shut off. " ACT TestQuestion ACT-Q22 question CT-22 Sample What is the size of the blood sample required to atart a test using the Precision Plus test strips? ACT Touch Touch Sample Please touch your answer. ACT Answer ACT-Q22A1 SpecialTest Sample 1 ml ACT Answer ACT-Q22A2 SpecialTest Sample .5 ml ACT Answer ACT-Q22A3 SpecialTest Sample 2 cc's ACT Answer ACT-Q22A4 SpecialTest Sample 5 microliters ACT Answer-R ACT-Q22A5 SpecialTest Sample 2.5 microliters (???r) ACT TestQuestion ACT-Q23 question CT-23 DataReview What button allows you to get to the Data Review option? ACT Touch Touch DataReview Please touch your answer. ACT Answer ACT-Q23A1 SpecialTest DataReview On/Off ACT Answer ACT-Q23A2 SpecialTest DataReview CLEAR ACT Answer ACT-Q23A3 SpecialTest DataReview ENTER ACT Answer-R ACT-Q23A5 SpecialTest DataReview MENU ACT Answer ACT-Q23A4 SpecialTest DataReview SCAN ACT TestQuestion ACT-Q24 question CT-24 DataPort "By placing the Precision XceedPro into the docking station, upload will begin automatically." ACT Touch Touch DataPort Please touch your answer. ACT Answer-R ACT-Q24A1 SpecialTest DataPort TRUE ACT Answer ACT-Q24A2 SpecialTest DataPort FALSE ACT TestQuestion ACT-Q25 question CT-25 DataPort What appears on the screen while a Precision XceedPro meter is uploading it's data? ACT Touch Touch DataPort Please touch your answer. ACT Answer ACT-Q25A1 SpecialTest DataPort Dock in Progress ACT Answer ACT-Q25A2 SpecialTest DataPort QC Due Now ACT Answer ACT-Q25A3 SpecialTest DataPort Patient Test or Control Test ACT Answer ACT-Q25A4 SpecialTest DataPort Remove Strip ACT Answer-R ACT-Q25A5 SpecialTest DataPort Please Wait Data Uploading ACT TestQuestion ACT-Q26 question CT-26 TS-PatientTest "What is the likely cause if you receive the error message, "" Strip Error Wet or Damaged""?" ACT Touch Touch TS-PatientTest Please touch your answer. ACT Answer-R ACT-Q26A5 SpecialTest TS-PatientTest The test strip is wet or has already been used. ACT Answer ACT-Q26A1 SpecialTest TS-PatientTest The test strip barcode did not scan properly. ACT Answer ACT-Q26A2 SpecialTest TS-PatientTest You have scanned or entered an expired strip. ACT Answer ACT-Q26A3 SpecialTest TS-PatientTest The test strip is the wrong kind for the meter. ACT TestQuestion ACT-Q27 question CT-27 PatientTest What is expected when the Precision XceedPro meter prompts you for the Patient Identification (ID)? ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q27A1 SpecialTest PatientTest Follow procedures to obtain a lab sample. ACT Answer ACT-Q27A2 SpecialTest PatientTest Scan the test strip barcode. ACT Answer ACT-Q27A3 SpecialTest PatientTest Enter in the Control solution lot number. ACT Answer ACT-Q27A4 SpecialTest PatientTest Press Enter to bypass the prompt. ACT Answer-R ACT-Q27A5 SpecialTest PatientTest Scan or Enter in the Patient Idenifier as described in your procedures. ACT TestQuestion ACT-Q28 question CT-28 PatientTest "What action is required when the Precision XceedPro Meter displays, Operator Certification Expires in 10 days." ACT Touch Touch PatientTest Please touch your answer. ACT Answer ACT-Q28A1 SpecialTest PatientTest You have scanned the test strip barcode by mistake. Scan a new test strip. ACT Answer ACT-Q28A2 SpecialTest PatientTest Try the test again later. ACT Answer ACT-Q28A3 SpecialTest PatientTest "No action is required, continue testing." ACT Answer-R ACT-Q28A5 SpecialTest PatientTest "Complete the test then, notify your supervisor/manager and obtain instructions for recertification." ACT TestQuestion ACT-Q29 question CT-29 Port What type of batteries does the Precision XceedPro Accept? ACT Touch Touch Port Please touch your answer. ACT Answer ACT-Q29A1 SpecialTest Port One 9 Volt ACT Answer ACT-Q29A2 SpecialTest Port Lithium Ion ACT Answer ACT-Q29A3 SpecialTest Port Nickel Cadmium ACT Answer ACT-Q29A4 SpecialTest Port Two C Batteries ACT Answer-R ACT-Q29A5 SpecialTest Port Two AA Batteries